The feasibility of using thermal strain imaging to regulate energy delivery during intracardiac radio-frequency ablation.
The feasibility of using thermal strain imaging to regulate energy delivery during intracardiac radio-frequency ablation.
复制标题
DOI:
10.1109/tuffc.2011.1960
复制
发表时间:
2011-07
期刊:
影响因子:
--
通讯作者:
O'Donnell M
中科院分区:
文献类型:
--
作者:
Seo CH;Stephens DN;Cannata J;Dentinger A;Lin F;Park S;Wildes D;Thomenius KE;Chen P;Nguyen T;de La Rama A;Jeong JS;Mahajan A;Shivkumar K;Nikoozadeh A;Oralkan O;Truong U;Sahn DJ;Khuri-Yakub PT;O'Donnell M
A method is introduced to monitor cardiac ablative therapy by examining slope changes in the thermal strain curve caused by speed of sound variations with temperature. The sound speed of water-bearing tissue such as cardiac muscle increases with temperature. However, at temperatures above about 50°C, there is no further increase in the sound speed and the temperature coefficient may become slightly negative. For ablation therapy, an irreversible injury to tissue and a complete heart block occurs in the range of 48 to 50°C for a short period in accordance with the well-known Arrhenius equation. Using these two properties, we propose a potential tool to detect the moment when tissue damage occurs by using the reduced slope in the thermal strain curve as a function of heating time. We have illustrated the feasibility of this method initially using porcine myocardium in vitro. The method was further demonstrated in vivo, using a specially equipped ablation tip and an 11-MHz microlinear intracardiac echocardiography (ICE) array mounted on the tip of a catheter. The thermal strain curves showed a plateau, strongly suggesting that the temperature reached at least 50°C.