NEUROPHYSIOLOGICAL INTRAOPERATIVE MONITORING OF THE GLOSSOPHARYNGEAL NERVE: TECHNICAL CASE REPORT

NEUROPHYSIOLOGICAL INTRAOPERATIVE MONITORING OF THE GLOSSOPHARYNGEAL NERVE: TECHNICAL CASE REPORT
复制标题

DOI:
10.1227/01.neu.0000316425.21752.e2
复制
发表时间:
2008-10
影响因子:
2.3
通讯作者:
A. Husain;D. Wright;Bret W. Stolp;A. Friedman;J. Keifer
A. Husain;D. Wright;Bret W. Stolp;A. Friedman;J. Keifer
中科院分区:
医学3区
文献类型:
--
作者:
A. Husain;D. Wright;Bret W. Stolp;A. Friedman;J. Keifer

文献摘要

被引文献

相似文献

OBJECTIVE Neurophysiological intraoperative monitoring of the glossopharyngeal nerve has been performed only with needle electrodes inserted into the pharyngeal muscles or soft palate. We describe a noninvasive method of monitoring this cranial nerve. METHODS A 30-year-old man who presented with headache, as well as speech and swallowing difficulty, underwent surgical resection of a right vagus nerve schwannoma. Neurophysiological intraoperative monitoring of multiple lower cranial nerves, including the glossopharyngeal and vagus nerves, was performed. RESULTS The glossopharyngeal nerve was monitored with an adhesive surface electrode mounted on the cuff of a laryngeal mask airway, and the vagus nerve was monitored with a similar electrode mounted on the endotracheal tube. Successful monitoring allowed separation of the glossopharyngeal nerve from the tumor, and there was no postoperative swallowing deficit. CONCLUSION Monitoring of the glossopharyngeal nerve with surface electrodes is possible and reliable, but it must be combined with vagus nerve monitoring.