Primary accelerometer calibration problems due to vibration exciters
Primary accelerometer calibration problems due to vibration exciters
复制标题
DOI:
10.1016/j.measurement.2009.05.002
复制
发表时间:
2009-11-01
期刊:
影响因子:
5.6
通讯作者:
Garcia, Guilherme A.
中科院分区:
文献类型:
--
作者:
Ripper, Gustavo P.;Dias, Ronaldo S.;Garcia, Guilherme A.
The purpose of this paper is to review and discuss the main problems that occur in primary accelerometer calibration due to the use of imperfect vibration exciters. These problems include low moving element stiffness, heating of the accelerometer under calibration, rocking and transverse motion, resonances, harmonic distortion and noise. Methods undertaken to minimize the effects of such problems are discussed and some calibration results are presented herein. (C) 2009 Elsevier Ltd. All rights reserved.