Hybrid phosphorescence and fluorescence native spectroscopy for breast cancer detection

Hybrid phosphorescence and fluorescence native spectroscopy for breast cancer detection
复制标题

DOI:
10.1117/1.2437139
复制
发表时间:
2007-01-01
影响因子:
3.5
通讯作者:
Alfano, R. R.
Alfano, R. R.
中科院分区:
医学3区
文献类型:
--
作者:
Alimova, Alexandra;Katz, A.;Alfano, R. R.

文献摘要

被引文献

相似文献

Fluorescence and phosphorescence measurements are performed on normal and malignant ex vivo human breast tissues using UV LED and xenon lamp excitation. Tryptophan (trp) phosphorescence intensity is higher in both normal glandular and adipose tissue when compared to malignant tissue. An algorithm based on the ratio of trp fluorescence intensity at 345 nm to phosphorescence intensity at 500 nm is successfully used to separate normal from malignant tissue types. Normal specimens consistently exhibited a low I-345/I-500 ratio (< 10), while for malignant specimens, the I-345/I-500 ratio is consistently high (> 15). The ratio analysis correlates well with histopathology. Intensity ratio maps with a spatial resolution of 0.5 mm are generated in which local regions of malignancy could be identified. (c) 2007 Society of Photo-Optical Instrumentation Engineers.