HIGH-RESOLUTION AND LOW-TEMPERATURE PHOTOEMISSION SPECTROSCOPY AT THE HISOR HELICAL-UNDULATOR BEAMLINE
HIGH-RESOLUTION AND LOW-TEMPERATURE PHOTOEMISSION SPECTROSCOPY AT THE HISOR HELICAL-UNDULATOR BEAMLINE
复制标题
DOI:
10.1142/s0218625x02002609
复制
发表时间:
2002-02
影响因子:
1.1
通讯作者:
M. Arita;K. Shimada;H. Namatame;M. Taniguchi
中科院分区:
文献类型:
--
作者:
M. Arita;K. Shimada;H. Namatame;M. Taniguchi
We report an experimental system dedicated for high-resolution and low-temperature photoemission spectroscopy installed on the helical-undulator beamline of a compact 700 MeV electron-storage ring (HiSOR) at Hiroshima University. A 3 m off-plane Eagle normal-incidence monochromator at the beamline covers the photon-energy range of hν = 4–40 eV. From the photoemission spectra near the Fermi edge of evaporated Au film, the total energy resolution was estimated to be 4.5 and 7.5 meV at hν = 7.1 and 22.6 eV, respectively. We have also performed photoemission measurements of Yb metal, which demonstrated highly bulk-sensitive spectra for the excitation by low energy photon. The high-resolution photoemission spectra of Y1 - xCaxTiO3 with x = 0.41 in the vicinity of the boundary between the metallic and insulating phases exhibited anomalous temperature dependence in comparison with those of normal metals.