Comparison of fast acquisition strategies in whole-heart four-dimensional flow cardiac MR: Two-center, 1.5 Tesla, phantom and in vivo validation study

Comparison of fast acquisition strategies in whole-heart four-dimensional flow cardiac MR: Two-center, 1.5 Tesla, phantom and in vivo validation study
复制标题

DOI:
10.1002/jmri.25746
复制
发表时间:
2018-01-01
影响因子:
4.4
通讯作者:
Plein, Sven
Plein, Sven
中科院分区:
医学2区
文献类型:
--
作者:
Garg, Pankaj;Westenberg, Jos J. M.;Plein, Sven

文献摘要

被引文献

相似文献

PurposeTo validate three widely-used acceleration methods in four-dimensional (4D) flow cardiac MR; segmented 4D-spoiled-gradient-echo (4D-SPGR), 4D-echo-planar-imaging (4D-EPI), and 4D-k-t Broad-use Linear Acquisition Speed-up Technique (4D-k-t BLAST).Materials and MethodsAcceleration methods were investigated in static/pulsatile phantoms and 25 volunteers on 1.5 Tesla MR systems. In phantoms, flow was quantified by 2D phase-contrast (PC), the three 4D flow methods and the time-beaker flow measurements. The later was used as the reference method. Peak velocity and flow assessment was done by means of all sequences. For peak velocity assessment 2D PC was used as the reference method. For flow assessment, consistency between mitral inflow and aortic outflow was investigated for all pulse-sequences. Visual grading of image quality/artifacts was performed on a four-point-scale (0=no artifacts; 3=nonevaluable).ResultsFor the pulsatile phantom experiments, the mean error for 2D PC=1.01.1%, 4D-SPGR=4.91.3%, 4D-EPI=7.61.3% and 4D-k-t BLAST=4.41.9%. In vivo, acquisition time was shortest for 4D-EPI (4D-EPI=8 +/- 2min versus 4D-SPGR=9 +/- 3min, P