Evaluation of a New 3D/2D Registration Criterion for Liver Radio-Frequencies Guided by Augmented Reality
Evaluation of a New 3D/2D Registration Criterion for Liver Radio-Frequencies Guided by Augmented Reality
复制标题
增强现实引导下肝脏射频新 3D/2D 配准标准的评估
DOI:
10.1007/3-540-45015-7_26
复制
发表时间:
2003
期刊:
影响因子:
--
通讯作者:
N. Ayache
中科院分区:
文献类型:
--
作者:
S. Nicolau;X. Pennec;L. Soler;N. Ayache
Our purpose in this article is to superimpose a 3D model of the liver, its vessels and tumors (reconstructed from CT images) on external video images of the patient for hepatic surgery guidance. The main constraints are the robustness, the accuracy and the computation time. Because of the absence of visible anatomical landmarks and of the “ cylindrical” shape of the upper abdomen, we used some radio-opaque fiducials. The classical least-squares method assuming that there is no noise on the 3D point positions, we designed a new Maximum Likelihood approach to account for this existing noise and we show that it generalizes the classical approaches. Experiments on synthetic data provide evidences that our new criterion is up to 20% more accurate and much more robust, while keeping a computation time compatible with realtime at 20 to 40 Hz. Eventually, careful validation experiments on real data show that an accuracy of 2 mm can be achieved within the liver.