Implantable Cardioverter-Defibrillator Lead Failure in Children and Young Adults

Implantable Cardioverter-Defibrillator Lead Failure in Children and Young Adults
复制标题

DOI:
10.1016/j.jacc.2013.09.033
复制
发表时间:
2014-01-21
影响因子:
24
通讯作者:
Shah, Maully J.
Shah, Maully J.
中科院分区:
医学1区
文献类型:
--
作者:
Janson, Christopher M.;Patel, Akash R.;Shah, Maully J.

文献摘要

被引文献

相似文献

Objectives This study aimed to investigate the impact of lead diameter and design on implantable cardioverter-defibrillator (ICD) lead survival in children and young adults. Background Recent reports have called attention to high rates of lead failure in adults with small-diameter ICD leads, but data in the pediatric population is limited.Methods We reviewed lead performance in consecutive subjects