Single-use MOSFET radiation dosimeters for the quality assurance of megavoltage photon beams
Single-use MOSFET radiation dosimeters for the quality assurance of megavoltage photon beams
复制标题
DOI:
10.1088/0031-9155/51/5/006
复制
发表时间:
2006-03-07
影响因子:
3.5
通讯作者:
Beddar, AS
中科院分区:
文献类型:
--
作者:
Briere, TM;Lii, J;Beddar, AS
We investigated the applicability of single-use MOSFET detectors as quality-assurance devices. Using ten accelerators available at our institution, we performed output measurements in both water and solid phantoms under photon irradiation. The MOSFET detectors performed well within the manufacturer's specifications, with average deviations of 2.1% and 0.7% for the 6 and 18 MV beams, respectively. The strength of the detector's design, including its wireless set-up, factory calibration and direct read-out, makes the system an acceptable independent quality-assurance device for use in verifying machine output within an accuracy of +/- 5%. The MOSFET detectors provide a quick check of machine output, which can be efficacious in detecting gross errors in machine calibrations.