False positive in the measurement of thyroglobulin induced by rheumatoid factor

False positive in the measurement of thyroglobulin induced by rheumatoid factor
复制标题

DOI:
10.1016/j.cca.2015.04.039
复制
发表时间:
2015-07-20
影响因子:
5
通讯作者:
Abalovich, M.
Abalovich, M.
中科院分区:
医学3区
文献类型:
--
作者:
Astarita, G.;Gutierrez, S.;Abalovich, M.

文献摘要

被引文献

相似文献

Background: We report a case of interference in thyroglobulin (Tg) measurement in a woman with differentiated thyroid cancer and rheumatoid arthritis history. Due to discordant Tg in relation to TSH concentrations and negative images, we investigated possible interference in the measurement of Tg.Methods: During the follow-up we measured Tg by chemiluminescence (TgQL) using Immulite 2000 immunoassay system. To investigate possible interference in Tg measurement, we made serial dilutions, re-testing of Tg by an alternative method: electrochemiluminescence, Cobas 6000 analyzer (TgEQL), recovery test of Tg and polyethylene glycol (PEG) 6000 precipitation.Results: During the patient follow-up, the TgQL ranged between