3D Printed Microfluidic Devices for Microchip Electrophoresis of Preterm Birth Biomarkers
3D Printed Microfluidic Devices for Microchip Electrophoresis of Preterm Birth Biomarkers
复制标题
DOI:
10.1021/acs.analchem.9b01395
复制
发表时间:
2019-06-04
影响因子:
7.4
通讯作者:
Woolley, Adam T.
中科院分区:
文献类型:
--
作者:
Beauchamp, Michael J.;Nielsen, Anna V.;Woolley, Adam T.
This work demonstrates for the first time the creation of microchip electrophoresis devices with similar to 50 mu m cross-sectional dimensions by stereolithographic 3D printing and their application in the analysis of medically significant biomarkers related to risk for preterm birth (PTB). We determined that device current was linear with applied potential up to 800 V (620 V/cm). We optimized device and separation conditions using fluorescently labeled amino acids as a model system and compared the performance in our 3D printed microfluidic devices to that in other device materials commonly used for microchip electrophoresis analysis. We demonstrated for the first time microchip electrophoresis in a 3D printed device of three PTB biomarkers, including peptides and a protein, with suitable separation characteristics. Limits of detection for microchip electrophoresis in 3D printed microfluidic devices were also determined for PTB biomarkers to be in the high picomolar to low nanomolar range.