IMPEDANCE CARDIOGRAPHY AS A NONINVASIVE METHOD OF MONITORING CARDIAC FUNCTION AND OTHER PARAMETERS OF CARDIOVASCULAR SYSTEM
IMPEDANCE CARDIOGRAPHY AS A NONINVASIVE METHOD OF MONITORING CARDIAC FUNCTION AND OTHER PARAMETERS OF CARDIOVASCULAR SYSTEM
复制标题
DOI:
10.1111/j.1749-6632.1970.tb17735.x
复制
发表时间:
1970-01-01
影响因子:
5.2
通讯作者:
WITSOE, DA
中科院分区:
文献类型:
--
作者:
KUBICEK, WG;PATTERSON, RP;WITSOE, DA
FIGURE 2. A photograph of the tape-on electrodes in place. It is important to maintain good separation between electrodcs 1 and 2. This electrode configuration is used for obtaining 1) cardiac function data and 2) changes in total fluid in the chest by observing Z,(the total impedance between electrodes 2 and 3). The Z, measurement has been found to be a very sensitive indicator of the development or reversal of such conditions as pulmonary edema, pulmonary congestion, and pleural effusion. electrode paste for thoracic impedance measurements. The electrode is lightweight and comfortable and possesses adequate electrical characteristics for transthoracic and limb impedance measurements. The tape-on electrodes are positioned on the subject as shown in FIGURE 2. The electrodes are easily attached to the impedance cardiograph by means of alligator clips on the patient-connecting cables.