Operation and Performance of a Mass-Selective Cryogenic Linear Ion Trap.
Operation and Performance of a Mass-Selective Cryogenic Linear Ion Trap.
复制标题
DOI:
10.1007/s13361-018-2026-7
复制
发表时间:
2018-11
影响因子:
3.2
通讯作者:
Polfer NC
中科院分区:
文献类型:
--
作者:
Tesler LF;Cismesia AP;Bell MR;Bailey LS;Polfer NC
We report on the performance of a cryogenic 2D linear ion trap (cryoLIT) that is shown to be mass-selective in the temperature range of 17 – 295K. As the cryoLIT is cooled, the ejection voltages during the mass instability scan decrease, which results in an effective mass shift to lower m/z relative to room temperature. This is attributed to a decrease in trap radius caused by thermal contraction. Additionally, the cryoLIT generates reproducible mass spectra from day-to-day, and is capable of performing stored waveform inverse Fourier transform (SWIFT) mass isolation of fragile N2-tagged ions for the purpose of background-free infrared dissociation spectroscopy.