Detailing radio frequency heating induced by coronary stents: a 7.0 Tesla magnetic resonance study.
Detailing radio frequency heating induced by coronary stents: a 7.0 Tesla magnetic resonance study.
复制标题
DOI:
10.1371/journal.pone.0049963
复制
发表时间:
2012
期刊:
影响因子:
3.7
通讯作者:
Niendorf T
中科院分区:
文献类型:
--
作者:
Santoro D;Winter L;Müller A;Vogt J;Renz W;Ozerdem C;Grässl A;Tkachenko V;Schulz-Menger J;Niendorf T
The sensitivity gain of ultrahigh field Magnetic Resonance (UHF-MR) holds the promise to enhance spatial and temporal resolution. Such improvements could be beneficial for cardiovascular MR. However, intracoronary stents used for treatment of coronary artery disease are currently considered to be contra-indications for UHF-MR. The antenna effect induced by a stent together with RF wavelength shortening could increase local radiofrequency (RF) power deposition at 7.0 T and bears the potential to induce local heating, which might cause tissue damage. Realizing these constraints, this work examines RF heating effects of stents using electro-magnetic field (EMF) simulations and phantoms with properties that mimic myocardium. For this purpose, RF power deposition that exceeds the clinical limits was induced by a dedicated birdcage coil. Fiber optic probes and MR thermometry were applied for temperature monitoring using agarose phantoms containing copper tubes or coronary stents. The results demonstrate an agreement between RF heating induced temperature changes derived from EMF simulations versus MR thermometry. The birdcage coil tailored for RF heating was capable of irradiating power exceeding the specific-absorption rate (SAR) limits defined by the IEC guidelines by a factor of three. This setup afforded RF induced temperature changes up to +27 K in a reference phantom. The maximum extra temperature increase, induced by a copper tube or a coronary stent was less than 3 K. The coronary stents examined showed an RF heating behavior similar to a copper tube. Our results suggest that, if IEC guidelines for local/global SAR are followed, the extra RF heating induced in myocardial tissue by stents may not be significant versus the baseline heating induced by the energy deposited by a tailored cardiac transmit RF coil at 7.0 T, and may be smaller if not insignificant than the extra RF heating observed under the circumstances used in this study.
登录
查看更多内容
DOI:
10.1186/1532-429x-14-15
发表时间:
2012-02-06
期刊:
Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance
影响因子:
--
作者:
Dall'Armellina E;Piechnik SK;Ferreira VM;Si QL;Robson MD;Francis JM;Cuculi F;Kharbanda RK;Banning AP;Choudhury RP;Karamitsos TD;Neubauer S
通讯作者:
Neubauer S
影响因子:
0.9
作者:
Huang, Shuning;Sosnovik, David E.
通讯作者:
Sosnovik, David E.
影响因子:
2.1
作者:
Rawson, Nigel S. B.;Chu, Rong;Terres, Jorge Alfonso Ross
通讯作者:
Terres, Jorge Alfonso Ross
影响因子:
37.8
作者:
Carpenter JP;He T;Kirk P;Roughton M;Anderson LJ;de Noronha SV;Sheppard MN;Porter JB;Walker JM;Wood JC;Galanello R;Forni G;Catani G;Matta G;Fucharoen S;Fleming A;House MJ;Black G;Firmin DN;St Pierre TG;Pennell DJ
通讯作者:
Pennell DJ
影响因子:
3.3
作者:
Nordbeck, Peter;Fidler, Florian;Quick, Harald H.
通讯作者:
Quick, Harald H.