Influence of the variation of energy spectra with depth in the dosimetry of 192Ir using LiF TLD.
Influence of the variation of energy spectra with depth in the dosimetry of 192Ir using LiF TLD.
复制标题
使用 LiF TLD 进行 192Ir 剂量测定时能谱随深度变化的影响。
DOI:
10.1088/0031-9155/33/10/005
复制
发表时间:
1988
影响因子:
3.5
通讯作者:
Nath,R
中科院分区:
文献类型:
--
作者:
Meigooni,AS;Meli,JA;Nath,R
Absolute dose measurements around low activity seeds used in brachytherapy are difficult to perform with ionisation chambers. The physical size of the chamber does not allow good resolution close to the seeds and the ionisation current generated is very low. Small thermoluminescent dosimeters (TLDs) overcome these problems but the energy dependence of their response should be considered. In this work, dose in polystyrene was measured at several distances from the high activity 192 Ir source (370 GBq) of a remote afterloading device using an ionisation chamber and LiF TLD chips. These data show that over a range of 1-10 cm from the source the sensitivity of LiF varies by up to 8.5%. This is attributed to the higher response of LiF to the lower photon energies, and to the shift of the photon spectrum to lower energies with increasing depth. The sensitivity of LiF to 192 Ir was also calculated by weighting the energy-dependent response of LiF by the Monte Carlo calculated photon spectra. The calculations give a similar change in sensitivity with distance from the source.