Comparison of two laser Doppler flowmeters for the monitoring of dermal blood flow
Comparison of two laser Doppler flowmeters for the monitoring of dermal blood flow
复制标题
两种激光多普勒流量计监测真皮血流的比较
DOI:
10.1002/micr.1920040304
复制
发表时间:
1983
期刊:
影响因子:
2.1
通讯作者:
W. Shaw
中科院分区:
文献类型:
--
作者:
J. Fischer;P. Parker;W. Shaw
Laser Doppler flowmetry (LDF) is a noninvasive and continuous method of monitoring dermal blood flow in free flaps. There are two LDFs available for clinical application: the LD5000 Laser Doppler Perfusion Monitor and the Periflux Laser Doppler Flowmeter. Varying degrees of arterial (N = 46) and venous (N = 28) occlusion were induced in saphenous island flaps of the dog, while simultaneous measurements with the LD5000 and Periflux LDFs were made. The millivoltage output of both the LD5000 and the Periflux showed a high degree of correlation with arterial and venous occlusion and with each other.