Performance of bioelectrical impedance analysis compared to dual X-ray absorptiometry (DXA) in Veterans with COPD.
Performance of bioelectrical impedance analysis compared to dual X-ray absorptiometry (DXA) in Veterans with COPD.
复制标题
DOI:
10.1038/s41598-022-05887-4
复制
发表时间:
2022-02-04
影响因子:
4.6
通讯作者:
Wan ES
中科院分区:
文献类型:
--
作者:
Cruz Rivera PN;Goldstein RL;Polak M;Lazzari AA;Moy ML;Wan ES
We examined the performance of a commercially-available handheld bioimpedance (BIA) device relative to dual X-ray absorptiometry (DXA) to assess body composition differences among Veterans with chronic obstructive pulmonary disease (COPD). Body composition was measured using DXA and BIA (Omron HBF-306C) at a single time point. Correlations between BIA- and DXA-assessed percent fat, fat mass, and fat-free mass were analyzed using Spearman (ρ) and Lin Concordance Correlation Coefficients (ρc). Mean differences in fat mass were visualized using Bland–Altman plots. Subgroup analyses by obesity status (BMI < 30 versus ≥ 30) were performed. Among 50 participants (96% male; mean age: 69.5 ± 6.0 years), BIA-assessed fat mass was strongly correlated (ρ = 0.94) and demonstrate excellent concordance (ρc = 0.95, [95%CI: 0.93–0.98]) with DXA, with a mean difference of 2.7 ± 3.2 kg between BIA and DXA. Although Spearman correlations between BIA- and DXA-assessed percent fat and fat-free mass were strong (ρ = 0.8 and 0.91, respectively), concordance values were only moderate (ρc = 0.67 and 0.74, respectively). Significantly stronger correlations were observed for obese relative to non-obese subjects for total percent fat (ρobese = 0.85 versus ρnon-obese = 0.5) and fat mass (ρobese = 0.96 versus ρnon-obese = 0.84). A handheld BIA device demonstrated high concordance with DXA for fat mass and moderate concordance for total percent fat and fat-free mass. ClinicalTrials.gov: NCT02099799.
登录
查看更多内容
影响因子:
6.9
作者:
Rush, Elaine C.;Crowley, Jennifer;Luke, Amy
通讯作者:
Luke, Amy
影响因子:
4.6
作者:
Robinson, Stephanie A.;Cooper, J. Allen, Jr.;Moy, Marilyn L.
通讯作者:
Moy, Marilyn L.
影响因子:
3.6
作者:
Lerario, M. C.;Sachs, A.;Jardim, J. R.
通讯作者:
Jardim, J. R.
影响因子:
2.8
作者:
Lan, Chou-Chin;Su, Chiu-Ping;Wu, Yao-Kuang
通讯作者:
Wu, Yao-Kuang
影响因子:
9.6
作者:
de Blasio, Francesca;Scalfi, Luca;de Blasio, Francesco
通讯作者:
de Blasio, Francesco