Blood glucose measurement in vivo using hollow-fiber based, mid-infrared ATR probe with multi-reflection prism
Blood glucose measurement in vivo using hollow-fiber based, mid-infrared ATR probe with multi-reflection prism
复制标题
使用带有多反射棱镜的中空纤维中红外 ATR 探头进行体内血糖测量
DOI:
10.1117/12.2216652
复制
发表时间:
2016
影响因子:
3.2
通讯作者:
Y. Matsuura
中科院分区:
文献类型:
--
作者:
S. Kino;Suguru Omori;Y. Matsuura
An attenuated-total-reflection (ATR), mid-infrared spectroscopy system that consists of hollow optical fibers, a trapezoidal multi-reflection ATR prism, and a conventional FT-IR spectrometer has been developed to measure blood glucose levels. Owing to the low transmission loss and high flexibility of the hollow-optical fiber, the system can measure any sites of the human body where blood capillaries are close to the surface of mucosa, such as inner lips. Using a multi-reflection prism brought about higher sensitivity, and the flat and wide contact surface of the prism resulted in higher measurement reproducibility. The results of in-vivo measurement of human inner lips showed the feasibility of the proposed system, and the measurement errors were within 20%.