Design of a Swept-Source, Anatomical OCT System for Pediatric Bronchoscopy.

Design of a Swept-Source, Anatomical OCT System for Pediatric Bronchoscopy.
复制标题

DOI:
10.1117/12.2004226
复制
发表时间:
2013-03-20
期刊:
Proceedings of SPIE--the International Society for Optical Engineering
影响因子:
--
通讯作者:
Oldenburg AL
Oldenburg AL
中科院分区:
其他
文献类型:
--
作者:
Wijesundara KC;Iftimia NV;Oldenburg AL

文献摘要

参考文献

相似文献

We describe the design and performance of a long coherence length, swept-source anatomical OCT (aOCT) system for pediatric airway imaging. A fiber-optic catheter is designed to be accommodated by a small-bore bronchoscope, and is scanned distally in a helical scan pattern to provide aOCT during bronchoscopy. We discuss particular challenges associated with the need for large imaging range, low SNR roll-off, and small catheter diameter. We present 3-D visualizations of airway phantoms and discuss optimization of the airway surface geometry obtained by aOCT. Accurate reconstruction of airway geometry will enable predictive modeling of patients suffering from airway obstruction.
DOI: 10.1109/50.97637
发表时间: 1991-11-01
影响因子: 4.7
作者:
KOHLHAAS, A;FROMCHEN, C;BRINKMEYER, E
通讯作者: BRINKMEYER, E
DOI: 10.1364/ao.42.003038
发表时间: 2003-06-01
期刊: APPLIED OPTICS
影响因子: 1.9
作者:
Marks, DL;Oldenburg, AL;Boppart, SA
通讯作者: Boppart, SA
DOI: 10.1364/oe.19.011880
发表时间: 2011-06-06
期刊: Optics express
影响因子: 3.8
作者:
Liu G;Rubinstein M;Saidi A;Qi W;Foulad A;Wong B;Chen Z
通讯作者: Chen Z
DOI: 10.1117/1.429942
发表时间: 1999-01-01
影响因子: 3.5
作者:
Yung, KM;Lee, SL;Schmitt, JM
通讯作者: Schmitt, JM
DOI: 10.1164/rccm.200507-1148oc
发表时间: 2006-01-15
影响因子: 24.7
作者:
Armstrong, JJ;Leigh, MS;Eastwood, PR
通讯作者: Eastwood, PR