Development of an immunosensor for the determination of allergy antibody (IgE) in blood samples
Development of an immunosensor for the determination of allergy antibody (IgE) in blood samples
复制标题
开发用于测定血液样本中过敏抗体 (IgE) 的免疫传感器
DOI:
10.1016/s0003-2670(01)01137-0
复制
发表时间:
2001
影响因子:
6.2
通讯作者:
G. Guilbault
中科院分区:
文献类型:
--
作者:
M. Kreuzer;C. O’Sullivan;M. Pravda;G. Guilbault
The current work describes the optimisation of a system for measurement of allergy antibodies (IgE) in blood samples using disposable screen-printed carbon electrode technology and amperometric detection of p-aminophenol, produced by alkaline phosphatase, at +300mV versus Ag/AgCl. The system was primarily developed by enzyme-linked immunosorbent assay (ELISA) prior to analysis by screen-printed electrodes. The sensor incorporates the clinically relevant range of IgE in sensitised patients with a detection limit of 0.09ngml−1IgE. The system is simple, requiring only addition of analyte (sample) and analysis is complete in 30min. In addition, it is cost effective and was developed for use as a point of care device, capable of measuring IgE levels in a drop of blood.