Low-intensity Pulsed Ultrasound Stimulation Significantly Enhances the Promotion of Bone Formation Around Dental Implants
Low-intensity Pulsed Ultrasound Stimulation Significantly Enhances the Promotion of Bone Formation Around Dental Implants
复制标题
低强度脉冲超声刺激显着增强牙种植体周围骨形成的促进作用
DOI:
10.2485/jhtb.20.139
复制
发表时间:
2011
影响因子:
0.4
通讯作者:
Hideki Matsubara
中科院分区:
文献类型:
--
作者:
Y. Nakanishi;Pao‐Li Wang;M. Ochi;K. Nakanishi;Hideki Matsubara
The effect of low-intensity pulsed ultrasound stimulation (LIPUS) on bone formation around dental implants was studied histologically and mechanically after a dental implant was inserted into the femur of male Japanese white rabbits. LIPUS daily treatment started on the day after implant placement for 14 days. Histologically, new bone formation was observed around the implant. Implant removal torque values were significantly higher in LIPUS-treated group (44.0 - 58.8 N·cm) than the control group (32.0N·cm). Moreover, bone contact ratio was significantly higher in LIPUS-treated group than the control group. The results suggest that clinical application of LIPUS for dental implants may promote osseointegration.