SIMULATION TO TREATMENT ERROR IN COMPUTED TOMOGRAPHY BASED TREATMENT PLANNING

基于计算机断层扫描的治疗计划中治疗误差的模拟

基本信息

DOI:
--
发表时间:
2000
期刊:
影响因子:
--
通讯作者:
Hisao Ito
中科院分区:
文献类型:
--
作者: T. Uno;S. Yasuda;T. Aruga;Koichi Isobe;R. Hara;N. Machida;H. Kawakami;Hisao Ito研究方向: -- MeSH主题词: --
关键词: --
来源链接:pubmed详情页地址

文献摘要

Purpose: To measure simulation to treatment error (STE) in radiation therapy and to evaluate factors which influence STE in computed tomography (CT) simulation. Materials & Methods: Off-line treatment verification using conventional portal films and CT-simulated digitally reconstructed radiographs (DRRs) was performed to measure the treatment accuracy in radiation therapy. Isocenter displacement of the portal films in relation to the DRRs was measured as a simulation to treatment error (STE). Correlation between STE and several treatment factors were examined statistically for 77 patients. Results: Factors that significantly affected STE were treatment site and use of immobilization devices. The mean STE of the patients with head and neck cancer was 3.9•}2.7 mm whereas that of patients with thoracic or abdominal tumor was 6.4•}6.1 mm (p=0.028). The mean STE in patients with and without immobilization device was 3.7•}2.6 mm and 5.7•}4.7 mm, respectively (p=0.027). Distance between isocenter and reference point at the time of planning CT scan did not correlate with STE. None of other factors influenced the accuracy of isocenter position. Conclusion: Although the position of isocenter was not marked at the time of CT-simulation, STE was within an acceptable range. Distance between the isocenter and reference point at the time of planning CT scan had no impact on STE.
目的:测量放射治疗中模拟与治疗的误差(STE),并评估影响计算机断层扫描(CT)模拟中STE的因素。 材料与方法:使用常规射野片和CT模拟的数字重建放射影像(DRR)进行离线治疗验证,以测量放射治疗的治疗准确性。将射野片相对于DRR的等中心位移作为模拟与治疗的误差(STE)进行测量。对77例患者的STE与若干治疗因素之间的相关性进行统计学检验。 结果:显著影响STE的因素是治疗部位和固定装置的使用。头颈部癌症患者的平均STE为3.9±2.7毫米,而胸部或腹部肿瘤患者的平均STE为6.4±6.1毫米(p = 0.028)。使用和未使用固定装置的患者的平均STE分别为3.7±2.6毫米和5.7±4.7毫米(p = 0.027)。计划CT扫描时等中心与参考点之间的距离与STE无关。其他因素均未影响等中心位置的准确性。 结论:尽管在CT模拟时未标记等中心位置,但STE在可接受范围内。计划CT扫描时等中心与参考点之间的距离对STE没有影响。
参考文献(2)
被引文献(0)
Comparison of computer workstation with light box for detecting setup errors from portal images.
计算机工作站与灯箱的比较,用于从门户图像中检测设置错误。
DOI:
10.1016/s0360-3016(99)00050-4
发表时间:
1999
期刊:
International journal of radiation oncology, biology, physics
影响因子:
0
作者:
Boxwala,AA;Chaney,EL;Fritsch,DS;Raghavan,S;Coffey,CS;Major,SA;Muller,KE
通讯作者:
Muller,KE

数据更新时间:{{ references.updateTime }}

Hisao Ito
通讯地址:
--
所属机构:
--
电子邮件地址:
--
免责声明免责声明
1、猫眼课题宝专注于为科研工作者提供省时、高效的文献资源检索和预览服务;
2、网站中的文献信息均来自公开、合规、透明的互联网文献查询网站,可以通过页面中的“来源链接”跳转数据网站。
3、在猫眼课题宝点击“求助全文”按钮,发布文献应助需求时求助者需要支付50喵币作为应助成功后的答谢给应助者,发送到用助者账户中。若文献求助失败支付的50喵币将退还至求助者账户中。所支付的喵币仅作为答谢,而不是作为文献的“购买”费用,平台也不从中收取任何费用,
4、特别提醒用户通过求助获得的文献原文仅用户个人学习使用,不得用于商业用途,否则一切风险由用户本人承担;
5、本平台尊重知识产权,如果权利所有者认为平台内容侵犯了其合法权益,可以通过本平台提供的版权投诉渠道提出投诉。一经核实,我们将立即采取措施删除/下架/断链等措施。
我已知晓