Examination of the safety of the exclusive liquid crystal display monitor for diagnostic imaging and research on the test picture development for accuracy management.
Examination of the safety of the exclusive liquid crystal display monitor for diagnostic imaging and research on the test picture development for accuracy management.
批准号:
15390359
负责人:
ISHIGAKI Takeo
金额:
$5.44万
依托单位:
依托单位国家:
日本
项目类别:
Grant-in-Aid for Scientific Research (B)
财政年份:
2003
资助国家:
日本
项目状态:
已结题
起止时间:
2003 至 2005
中文摘要
液晶显示器的基本检查显示,IPS (In-plane Switching)模式和VA (vertical Aligned)模式最适合作为诊断成像的对象。虽然与CRT显示器相比,液晶显示器的亮度衰减时间要宽松得多,但当连续10000小时后,亮度下降到初始值的70-80%。早期亮度保持在监视器中,其中包括亮度稳定电路至少10,000小时。在对比度较低的情况下,可以用液晶显示器代替CRT显示器。室内照度的影响不显著,建议在明亮的照明下使用液晶显示器。即使与CRT显示器比较关于液晶显示器的诊断能力(1M, 2M, 3M, 5M)。也就是说,可以用CRT显示器代替液晶显示器进行诊断成像(乳房x光检查除外)。在使用3M液晶显示器的乳房x光片研究中,未见对不可触及的钙化病变的诊断能力下降。从这些研究结果可以得出结论,用于CRT显示器亮度退化判断的测试图片(临床图片和对比图)也适用于液晶显示器亮度退化判断。
英文摘要
Fundamental examination of a liquid crystal display monitor showed that IPS (In-plane Switching) mode and VA (Vertically Aligned) mode were most suitable as an object for diagnostic imaging. Although the time of luminosity degradation of a liquid crystal display monitor was far loose compared with the CRT monitor, when it became in continuation 10,000 hours, it turned out that luminosity falls to 70-80% of an initial value. Early luminosity is kept in the monitor in which the luminosity stabilization circuit is included for at least 10,000 hours. At the target of low contrast, a liquid crystal display can be used as substitution of a CRT monitor. The influence of indoor illumination was not significant and it was suggested also under bright lighting that a liquid crystal display can be used. even if compared with a CRT monitor about the diagnostic ability of a liquid crystal display monitor (1M, 2M, 3M, and 5M). That is, a liquid crystal display monitor can be replaced with a CRT monitor for diagnostic imaging (except for mammography). In the research of a mammo gram using 3M liquid crystal display monitor, the fall of the diagnostic ability of a non-palpable calcification lesion was not seen. From these research results, it is concluded that the test picture for the luminosity degradation judging of a CRT monitor (a clinical picture and contrast chart) would be applicable also to the luminosity degradation judging of a liquid crystal display monitor.
期刊论文(2)
专著(0)
科研奖励(0)
会议论文
画像観察モニタに関する諸問題
图像观察监视器的相关问题
DOI:
--
发表时间:
2005
期刊:
映像情報メディカル 37巻4号
影响因子:
--
作者:
[石垣武男]
通讯作者:
石垣武男
Theoretical analysis of diagnostic process of medical image and optimization of CRT image processing
-
批准号:09307016
-
项目类别:Grant-in-Aid for Scientific Research (A)
-
资助金额:$12.29万
-
财政年份:1997
-
负责人:ISHIGAKI Takeo
-
依托单位:
Generation of the workstation for image interpretation based on observer's diagnosis process
-
批准号:07557238
-
项目类别:Grant-in-Aid for Scientific Research (A)
-
资助金额:$5.31万
-
财政年份:1995
-
负责人:ISHIGAKI Takeo
-
依托单位:
Assessment of the effectiveness of cancer therapy using PET.
-
批准号:06670914
-
项目类别:Grant-in-Aid for Scientific Research (C)
-
资助金额:$1.34万
-
财政年份:1994
-
负责人:ISHIGAKI Takeo
-
依托单位: