SEGMENTAL PULSATILE PRESSURE CUFF & ANALYZER DEVELOPMENT
SEGMENTAL PULSATILE PRESSURE CUFF & ANALYZER DEVELOPMENT
批准号:
3508940
负责人:
HAYATI BALKANLI
金额:
$19.41万
依托单位国家:
美国
项目类别:
财政年份:
1986
资助国家:
美国
项目状态:
已结题
起止时间:
1986-09-30 至 1988-08-31
中文摘要
点击翻译按钮获取中文摘要
英文摘要
The proposed project will generate an accurate pressure analyzer and a
segmental penile cuff for sensing penile blood pressure pulses. Based on
the results of the feasibility work undertaken, the specifications of the
analyzer will encompass an overall accuracy of 0.05%, a resolution of
0.01mm Hg pressure icrement, a frequency and from zero to ten Hz (at 3 db),
and a very low, tolerable leakage in its pneumatic system with no leakage
through the transducer. Even though the target application goal is to
diagnose and evaluate male impotence, the analyzer can satisfy a wide
variety of applications. The analyzer with a cuff is readily applicable to
cardiovascular research. With a suitable pressure pickup sensor the
analyzer can be used for measurement of intracranial pressure in trauma
centers and hospitals. The immediate application market of the proposed
cuff and analyzer system is in sleep laboratories and in urology clinics.
The next targeted market of the analyzer is trauma centers. In order to
achieve the above mentioned accuracy the analyzer will be equipped with a
built-in electronic calibration chart which will correct the
non-linearities of the transfer function of the pressure transducer. The
hermeticity of the transducer will be checked by a helium leak detector
before installation into the system. Since the analyzer will provide both
analog and digtal data output, it will be compatible with both analog
recorders and computers. For the applications requiring wider frequency
bandwidth the transducer can be improved. Its computer compatibility can
facilitate real time analysis.