- 关键词:X raybioengineering /biomedical engineeringbiomedical equipment developmentbiomedical equipment safetybreast neoplasm /cancer diagnosisclinical researchclinical trial phase Icomputed axial tomographycomputer assisted diagnosiscomputer simulationcomputer system design /evaluationdiagnosis design /evaluationdiagnosis quality /standardearly diagnosisfemalehuman subjectimage enhancementmathematicspatient oriented researchradiation dosagewomen&aposs health
项目摘要
DESCRIPTION (provided by applicant): Breast cancer is a disease with high
incidence in the U.S. and elsewhere, and population-level methods of fighting
this disease are aimed primarily on screening, using mammography for early
detection. The median size of breast cancer found using mammography is
approximately 11 mm. Based on extensive preliminary studies involving computer
simulations, physical measurements, and cadaver breast imaging, we have found
that breast CT may be able to routinely detect much smaller breast tumors, in
the 3 to 5 mm range. Importantly, the radiation dose of breast CT performed at
80 kVp was found in detailed studies to be comparable to that of mammography.
It is not possible to image the breast alone on a live woman using a clinical
CT scanner. Therefore, in this Bioengineering Research Partnership proposal, we
have teamed with scientists from around the country to design, build, and test
a CT scanner designed to image the breast. A team comprised of medical
physicists, physicians, mechanical and electrical engineers, and breast cancer
advocates will collaborate on the design of the breast CT scanner. Cone beam
flat panel technology will be used to produce a scanner capable of 10 second
breast scanning, and the scanner development will also include a breast
immobilization system (acrylic cylinders), a breast CT table, a fast
reconstruction computer, and a computer workstation customized for efficient
viewing breast CT images. The scanner will be built, tested, and optimized at
UC Davis over a period of 3 years involving 9 specific aims. After the breast
CT scanner is tested in a brief phase I trial (2 specific aims), it will be
moved to the breast imaging clinic for a phase II trial where approximately 120
women will be imaged (4 specific aims). This phase II trial will evaluate the
efficacy of breast CT for the early detection of breast cancer in a group of
women likely to have breast cancer (BIRADS 4 & 5). Additionally, the breast
image data will be studied for its utility in automating the analysis of the
normal breast architecture, and for computerized cancer detection. In year 5 of
the proposed research, two specific aims utilize the breast CT data and
corresponding mammography images (on -240 breasts) to evaluate the ideal
observer performance and human (mammographer) detection performance attributes
of the breast CT scanner. At the end of the proposed research involving 17
specific aims, the potential of breast CT will have been evaluated both
qualitatively and quantitatively. A tested, high quality prototype breast CT
scanner would be ready to be enlisted in a phase III trial (beyond the scope of
this proposed research), if further testing is warranted. Performance data
acquired in the present study would allow the proper design (power, etc.) of a
phase III trial. If breast CT lives up to its enormous potential based on
initial imaging, breast cancer would be detectable far before metastases occurs
- for example, a 3 mm tumor contains only 2 percent of the cell count of an 11
mm lesion, and a 5 mm lesion contains only 9 percent of the cell count. Based
on a 100 day volume doubling time, detection of a 5 mm lesion would lead to
0.93 year earlier detection, and routine detection of 3 mm lesions would result
in 1.5 year earlier detection over mammography. Surgical removal of early
cancers will effectively result in cure for the majority of women screened
using this technology. While breast CT would probably improve cancer detection
in all women, some women may have risk factors (dense breasts, genetic markers,
etc.) that particularly warrant screening using breast CT. The Phase II trial
will shed more light on this issue.
描述(申请人提供):乳腺癌是一种高风险的疾病。
美国和其他地方的发病率,以及人口层面的战斗方法
本病主要用于筛查,早期使用乳房X光检查
侦测。使用乳房X光摄影发现的乳腺癌的中位数是
大约11毫米。基于涉及计算机的广泛的初步研究
模拟、物理测量和身体乳房成像,我们发现
乳腺CT也许能够常规地检测出更小的乳腺肿瘤,在
3到5毫米的范围。重要的是,乳腺CT的辐射剂量在
在详细的研究中发现80kVp与乳房X光摄影的结果相当。
在活体妇女身上单独成像乳房是不可能的,使用临床
CT扫描仪。因此,在这项生物工程研究伙伴关系提案中,我们
已经与来自全国各地的科学家合作设计、建造和测试
用来对乳房进行成像的CT扫描仪。由医疗队组成的团队
物理学家、内科医生、机械和电气工程师以及乳腺癌
倡导者将合作设计乳腺CT扫描仪。锥形梁
平板技术将被用于生产一种能够10秒的扫描仪
乳房扫描,而扫描仪的开发还将包括乳房
固定系统(亚克力圆柱体),乳房CT床,FAST
重建计算机和为高效定制的计算机工作站
查看乳腺CT图像。扫描仪将在以下位置进行构建、测试和优化
加州大学戴维斯分校为期3年,涉及9个具体目标。乳房过后
CT扫描仪在简短的第一阶段试验中进行测试(2个具体目标),它将
转到乳房成像诊所进行II期试验,大约有120人
女性将被成像(4个具体目标)。这项第二阶段试验将评估
乳腺CT在早期发现乳腺癌中的作用
可能患有乳腺癌的妇女(BIRADS 4和5)。此外,乳房
我们将研究图像数据在自动分析
正常的乳房结构,以及计算机化的癌症检测。年第5年
拟议的研究有两个具体目标,利用乳房CT数据和
相应的乳房X光检查图像(对-240个乳房)进行评估
观察者性能和人类(乳房X光摄影者)检测性能属性
乳房CT扫描仪。在拟议的研究结束时,涉及17
特定的目标,乳腺CT的潜力将被评估为
定性的和定量的。经过测试的高质量乳腺CT原型
扫描仪将准备好进入第三阶段试验(超出
这项拟议的研究),如果进一步的测试是必要的。性能数据
在本研究中获得的将允许适当的设计(电源等)一种
第三阶段试验。如果乳腺CT发挥其巨大的潜力,基于
最初的成像,乳腺癌在转移发生之前就可以被检测到
-例如,一个3 mm的肿瘤只包含11个肿瘤的2%的细胞
5 mm的病变仅含9%的细胞计数。基座
在100天的体积倍增时间内,检测到5 mm的病变将导致
0.93年前检测,并常规检测3 mm病变
在1.5年前检测到的乳房X光检查。早期手术切除
癌症将有效地治愈大多数接受筛查的妇女
使用这项技术。而乳腺CT可能会提高癌症的检测能力
在所有女性中,一些女性可能有风险因素(丰满的乳房,遗传标记,
等)尤其需要使用乳房CT进行筛查。第二阶段试验
将在这个问题上提供更多的线索。
项目成果
期刊论文数量(1)
专著数量(0)
科研奖励数量(0)
会议论文数量(0)
专利数量(0)
Dedicated breast computed tomography: volume image denoising via a partial-diffusion equation based technique.
- DOI:10.1118/1.2903436
- 发表时间:2008-05
- 期刊:
- 影响因子:3.8
- 作者:J. Xia;J. Lo;Kai Yang;C. Floyd;J. Boone
- 通讯作者:J. Xia;J. Lo;Kai Yang;C. Floyd;J. Boone
{{
item.title }}
{{ item.translation_title }}
- DOI:
{{ item.doi }} - 发表时间:
{{ item.publish_year }} - 期刊:
- 影响因子:{{ item.factor }}
- 作者:
{{ item.authors }} - 通讯作者:
{{ item.author }}
数据更新时间:{{ journalArticles.updateTime }}
{{ item.title }}
- 作者:
{{ item.author }}
数据更新时间:{{ monograph.updateTime }}
{{ item.title }}
- 作者:
{{ item.author }}
数据更新时间:{{ sciAawards.updateTime }}
{{ item.title }}
- 作者:
{{ item.author }}
数据更新时间:{{ conferencePapers.updateTime }}
{{ item.title }}
- 作者:
{{ item.author }}
数据更新时间:{{ patent.updateTime }}
JOHN M BOONE其他文献
JOHN M BOONE的其他文献
{{
item.title }}
{{ item.translation_title }}
- DOI:
{{ item.doi }} - 发表时间:
{{ item.publish_year }} - 期刊:
- 影响因子:{{ item.factor }}
- 作者:
{{ item.authors }} - 通讯作者:
{{ item.author }}
{{ truncateString('JOHN M BOONE', 18)}}的其他基金
Hybrid Nuclear/X-ray Projection System for Mouse Imaging
用于小鼠成像的混合核/X 射线投影系统
- 批准号:
6951606 - 财政年份:2004
- 资助金额:
$ 85.35万 - 项目类别:
Hybrid Nuclear/X-ray Projection System for Mouse Imaging
用于小鼠成像的混合核/X 射线投影系统
- 批准号:
6848595 - 财政年份:2004
- 资助金额:
$ 85.35万 - 项目类别:
Breast CT scanner for earlier cancer detection
用于早期癌症检测的乳房 CT 扫描仪
- 批准号:
6933181 - 财政年份:2002
- 资助金额:
$ 85.35万 - 项目类别:
Breast CT scanner for earlier cancer detection
用于早期癌症检测的乳房 CT 扫描仪
- 批准号:
8060548 - 财政年份:2002
- 资助金额:
$ 85.35万 - 项目类别:
Breast CT scanner for earlier cancer detection
用于早期癌症检测的乳房 CT 扫描仪
- 批准号:
6789982 - 财政年份:2002
- 资助金额:
$ 85.35万 - 项目类别:
Breast CT scanner for earlier cancer detection
用于早期癌症检测的乳房 CT 扫描仪
- 批准号:
7342661 - 财政年份:2002
- 资助金额:
$ 85.35万 - 项目类别:
Breast CT scanner for earlier cancer detection
用于早期癌症检测的乳房 CT 扫描仪
- 批准号:
8269559 - 财政年份:2002
- 资助金额:
$ 85.35万 - 项目类别:














{{item.name}}会员