A Hybrid Continuous Glucose and Lactate Monitor with Improved Signal Reliability
A Hybrid Continuous Glucose and Lactate Monitor with Improved Signal Reliability
批准号:
10597237
负责人:
Jared Rylan Tangney
金额:
$78.64万
依托单位:
依托单位国家:
美国
项目类别:
财政年份:
2021
资助国家:
美国
项目状态:
已结题
起止时间:
2021-09-01 至 2024-06-30
关键词:
AppleArchivesBiosensorBlood GlucoseBluetoothBuffersBusinessesCharacteristicsChemistryClinicalClinical ResearchCommunitiesContinuous Glucose MonitorCustomData DisplayDatabasesDermalDermisDetectionDevelopmentDevicesDiabetes MellitusDiameterDimensionsDoseEffectivenessElementsEnsureEvaluationFingersGenerationsGlucoseGlucose Plasma ConcentrationHeightHormonesHourHybridsHypoglycemiaIn VitroIndividualInsulinInsulin-Dependent Diabetes MellitusIntercellular FluidLeadMeasurementMeasuresMonitorNatureNeedlesNoiseOxidasesPain FreePatientsPerformancePhasePhosphate BufferPhysical activityPhysiologic pulsePhysiologicalPropertyReportingResearchResolutionSafetySamplingSecureSignal TransductionSkinSterilitySterilizationSurfaceSystemSystems DevelopmentTechnologyTimeTransducersTranslatingValidationWorkbattery lifebiomaterial compatibilitycirculating biomarkerscloud basedcommercializationcostdesigndetection limitdiabetes managementimprovedindividual variationinsightintegrated circuitminiaturizemobile applicationmonitoring devicenext generationnovelparallelizationrate of changeresponsesensorwireless
中文摘要
点击翻译按钮获取中文摘要
英文摘要
PROJECT SUMMARY
Over the past decade, the diabetes community has benefited enormously from the successful commercialization
of multiple generations of continuous glucose monitors (CGMs). These devices are able to report readings of
glucose in the interstitial fluid (ISF) to users on a near real-time basis. Over the few years, driven primarily by
demands for improved glycemic management in insulin-dependent diabetes, the accuracy and reliability of CGM
devices made further advances leading to FDA approval of their use in determining insulin dosing without the
need for confirmatory fingersticks. The latest generation of CGM devices are routinely used to provide input to
automated insulin delivery (AID) systems. However, despite these impressive improvements in CGM technology,
CGMs today remain limited to the measurement of a single analyte and lack the ability to take redundant
measurements. It is increasingly recognized that further advances in glycemic management will likely require
the addition of other circulating biomarkers to elucidate individual variability in glycemic response, especially to
physical activity. Measurement of freely circulating lactate can augment established glucose measurements as
lactate has been found to shift glycemic thresholds for hypoglycemia in individuals with T1D to lower plasma
glucose concentrations [33]. This project thus aims to provide a novel CGM platform capable of accurately
measuring both glucose and lactate. We believe that a combined glucose and lactate continuous monitoring
system will enhance the safety and efficacy of the next generation of AID systems. Biolinq has developed a
body-worn patch containing a microneedle array with multiplexed sensing elements. Clinical studies to date have
used a microneedle array consisting of individually-addressable microneedles inserted into the reticular dermis
– approximately 200 to 350 microns below the surface of the skin. Previous work has shown a high level of
correlation between glucose in the dermal interstitial fluid and blood glucose [57]. Our clinical studies have
confirmed the high level of correlation reported and low lag time previously established for glucose in the dermal
ISF [36]; we believe this is a unique property of sensing in the dermis and can be extended to other metabolites.
In accordance with the project plan, the research effort seeks to attain the following objectives: (1) development
and refinement of a lactate-selective sensing chemistry and its co-location with Biolinq’s clinically-validated
glucose-selective chemistry, (2) implementation of redundant measures of glucose and lactate in a microneedle
array sensor platform featuring individually-addressable sensing elements to provide accurate and reliable
measurements in the dermal ISF, and (3) design of an end-to-end body-worn sensing system containing a
miniaturized, low-power application-specific integrated circuit with Bluetooth wireless capability. The technology
and business objectives outlined in this proposal will have a direct impact on open- and closed-loop insulin
delivery and should meet current needs for an accurate, non-invasive continuous glucose and lactate monitor
with broad applicability owing to its enhanced reliability, accuracy, simple use profile, and low-cost nature.