Electron FLASH Delivery at Treatment Room Isocenter for Efficient Reversible Conversion of a Clinical LINAC.
Electron FLASH Delivery at Treatment Room Isocenter for Efficient Reversible Conversion of a Clinical LINAC.
复制标题
DOI:
10.1016/j.ijrobp.2021.01.011
复制
发表时间:
2021-07-01
影响因子:
7
通讯作者:
Gladstone, David J.
中科院分区:
文献类型:
--
作者:
Rahman, Mahbubur;Ashraf, Ramish;Zhang, Rongxiao;Bruza, Petr;Dexter, Chad A.;Thompson, Lawrence;Cao, Xu;Williams, Benjamin B.;Hoopes, P. Jack;Pogue, Brian W.;Gladstone, David J.
In this study, procedures were developed to achieve efficient reversible conversion of a clinical linear accelerator (LINAC) and deliver ultra-high dose-rate (UHDR) electron or conventional beams to the treatment room isocenter for FLASH radiotherapy. The LINAC was converted to deliver UHDR beam within 20 minutes by retracting the x-ray target from the beam’s path, positioning the carousel on an empty port, and selecting 10 MV photon beam energy in the treatment console. Dose-rate, surface and depth dose profiles were measured in solid water phantom at different field sizes with Gafchromic film and optically stimulated luminescent dosimeter (OSLD). A pulse controller counted the pulses via scattered radiation signal and gated the delivery for preset pulse count. A fast photomultiplier tube-based Cherenkov detector measured per pulse beam output at 2 ns sampling rate. After conversion back to clinical mode, conventional beam output, flatness, symmetry, field size and energy were measured for all clinically commissioned energies. The surface average dose rate at the isocenter for a 1 cm in diameter, 1.5 in diameter circular fields, and jaws wide open field was 238±5, 262±5, and 290±5 Gy/s, respectively. The radial symmetry of the beams was within 2.4%, 0.5%, and 0.2% respectively. The doses from simultaneous irradiation of film and OSLD were within 1%. The PMT showed the LINAC required ramp up time in the first 4–6 pulses before the output stabilized and its stability was within 3%. 10 MeV UHDR beams were achieved at the isocenter of the treatment room. The beam output was reproducible but requires further investigation of the ramp up time, equivalent to ~1 Gy, requiring dose monitoring. The UHDR beam can irradiate both small and large subjects to investigate potential FLASH radiobiological effects in minimally modified clinical settings and dose-rates can be further increased by reducing the source to surface distance.
登录
查看更多内容
影响因子:
5.7
作者:
Montay-Gruel, Pierre;Bouchet, Audrey;Vozenin, Marie-Catherine
通讯作者:
Vozenin, Marie-Catherine
影响因子:
3.8
作者:
Lewis, David;Micke, Andre;Chan, Maria F.
通讯作者:
Chan, Maria F.
影响因子:
3.8
作者:
Lewis, David F.;Chan, Maria F.
通讯作者:
Chan, Maria F.
影响因子:
3.8
作者:
ARCOVITO, G;PIERMATTEI, A;BASSI, FA
通讯作者:
BASSI, FA
影响因子:
3.5
作者:
Khachonkham, Suphalak;Dreindl, Ralf;Kuess, Peter
通讯作者:
Kuess, Peter