Sensitivity of portable low-field magnetic resonance imaging for multiple sclerosis lesions.
Sensitivity of portable low-field magnetic resonance imaging for multiple sclerosis lesions.
复制标题
DOI:
10.1016/j.nicl.2022.103101
复制
发表时间:
2022
影响因子:
4.2
通讯作者:
Stein, Joel M.
中科院分区:
文献类型:
--
作者:
Arnold, Campbell;Tu, Danni;Okar, Serhat, V;Nair, Govind;By, Samantha;Kawatra, Karan D.;Robert-Fitzgerald, Timothy E.;Desiderio, Lisa M.;Schindler, Matthew K.;Shinohara, Russell T.;Reich, Daniel S.;Stein, Joel M.
Paired, same-day, 3T and 64mT MRI studies were analyzed in 33 MS patients. 64mT MRI showed 94% sensitivity for detecting any lesions in 3T confirmed cases. The diameter of the smallest detected lesion was larger at 64mT compared to 3T. Total lesion volume estimates were strongly correlated between 3T and 64mT scans. Portable low-field MRI detects white matter lesions, but smaller lesions may be missed. Magnetic resonance imaging (MRI) is a fundamental tool in the diagnosis and management of neurological diseases such as multiple sclerosis (MS). New portable, low-field strength, MRI scanners could potentially lower financial and technical barriers to neuroimaging and reach underserved or disabled populations, but the sensitivity of these devices for MS lesions is unknown. We sought to determine if white matter lesions can be detected on a portable 64mT scanner, compare automated lesion segmentations and total lesion volume between paired 3T and 64mT scans, identify features that contribute to lesion detection accuracy, and explore super-resolution imaging at low-field. In this prospective, cross-sectional study, same-day brain MRI (FLAIR, T1w, and T2w) scans were collected from 36 adults (32 women; mean age, 50 ± 14 years) with known or suspected MS using Siemens 3T (FLAIR: 1 mm isotropic, T1w: 1 mm isotropic, and T2w: 0.34–0.5 × 0.34–0.5 × 3–5 mm) and Hyperfine 64mT (FLAIR: 1.6 × 1.6 × 5 mm, T1w: 1.5 × 1.5 × 5 mm, and T2w: 1.5 × 1.5 × 5 mm) scanners at two centers. Images were reviewed by neuroradiologists. MS lesions were measured manually and segmented using an automated algorithm. Statistical analyses assessed accuracy and variability of segmentations across scanners and systematic scanner biases in automated volumetric measurements. Lesions were identified on 64mT scans in 94% (31/33) of patients with confirmed MS. The average smallest lesions manually detected were 5.7 ± 1.3 mm in maximum diameter at 64mT vs 2.1 ± 0.6 mm at 3T, approaching the spatial resolution of the respective scanner sequences (3T: 1 mm, 64mT: 5 mm slice thickness). Automated lesion volume estimates were highly correlated between 3T and 64mT scans (r = 0.89, p < 0.001). Bland-Altman analysis identified bias in 64mT segmentations (mean = 1.6 ml, standard error = 5.2 ml, limits of agreement = –19.0–15.9 ml), which over-estimated low lesion volume and under-estimated high volume (r = 0.74, p < 0.001). Visual inspection revealed over-segmentation was driven venous hyperintensities on 64mT T2-FLAIR. Lesion size drove segmentation accuracy, with 93% of lesions > 1.0 ml and all lesions > 1.5 ml being detected. Using multi-acquisition volume averaging, we were able to generate 1.6 mm isotropic images on the 64mT device. Overall, our results demonstrate that in established MS, a portable 64mT MRI scanner can identify white matter lesions, and that automated estimates of total lesion volume correlate with measurements from 3T scans.
登录
查看更多内容
影响因子:
5.7
作者:
Jovicich J;Czanner S;Han X;Salat D;van der Kouwe A;Quinn B;Pacheco J;Albert M;Killiany R;Blacker D;Maguire P;Rosas D;Makris N;Gollub R;Dale A;Dickerson BC;Fischl B
通讯作者:
Fischl B
影响因子:
16.6
作者:
Mazurek MH;Cahn BA;Yuen MM;Prabhat AM;Chavva IR;Shah JT;Crawford AL;Welch EB;Rothberg J;Sacolick L;Poole M;Wira C;Matouk CC;Ward A;Timario N;Leasure A;Beekman R;Peng TJ;Witsch J;Antonios JP;Falcone GJ;Gobeske KT;Petersen N;Schindler J;Sansing L;Gilmore EJ;Hwang DY;Kim JA;Malhotra A;Sze G;Rosen MS;Kimberly WT;Sheth KN
通讯作者:
Sheth KN
影响因子:
4.7
作者:
Deoni SCL;D'Sa V;Volpe A;Beauchemin J;Croff JM;Elliott AJ;Pini N;Lucchini M;Fifer WP
通讯作者:
Fifer WP
影响因子:
9.9
作者:
Hagens MHJ;Burggraaff J;Kilsdonk ID;de Vos ML;Cawley N;Sbardella E;Andelova M;Amann M;Lieb JM;Pantano P;Lissenberg-Witte BI;Killestein J;Oreja-Guevara C;Ciccarelli O;Gasperini C;Lukas C;Wattjes MP;Barkhof F;MAGNIMS Study Group
通讯作者:
MAGNIMS Study Group
影响因子:
3.5
作者:
Mattay, R. R.;Davtyan, K.;Mamourian, A. C.
通讯作者:
Mamourian, A. C.