Label-free porous silicon immunosensor for broad detection of opiates in a blind clinical study and results comparison to commercial analytical chemistry techniques.

Label-free porous silicon immunosensor for broad detection of opiates in a blind clinical study and results comparison to commercial analytical chemistry techniques.
复制标题

DOI:
10.1021/ac101804s
复制
发表时间:
2010-12-01
影响因子:
7.4
通讯作者:
DeLouise, Lisa A.
DeLouise, Lisa A.
中科院分区:
化学1区
文献类型:
--
作者:
Bonanno, Lisa M.;Kwong, Tai C.;DeLouise, Lisa A.

文献摘要

参考文献

被引文献

相似文献

在这项工作中,我们首次评估了一种无标记多孔硅(PSI)免疫传感器试验在一项盲目临床研究中的性能,该试验旨在筛选各种阿片类药物的真实患者尿样。PSI阿片免疫传感器与液质联用(LC-MS/MS)检测结果的符合率为96%。此外,成功检测到一种常见的滥用阿片类药物羟考酮,使PSI阿片传感器与LC-MS/MS之间的定性符合率达到100%。相比之下,商用广谱阿片免疫分析技术(CEDIA®)与LC-MS/MS的定性符合率达到65%。对添加了测试分析物的空白尿样进行了精密度、准确性和回收率等重要性能属性的评估。作为模型阿片靶标的吗啡检测的变异系数在运行内和日间均为9%,超过300ngmlPSI 1的截止限值。这项研究验证了无标记−阿片免疫传感器在真实患者样本中的分析筛选能力,并首次半定量展示了该技术的成功临床应用。这些结果推动了PSI技术的未来发展,以降低诊断测试的复杂性和成本,特别是在医疗保健环境中。
In this work we evaluate for the first time the performance of a label-free porous silicon (PSi) immunosensor assay in a blind clinical study designed to screen authentic patient urine specimens for a broad range of opiates. The PSi opiate immunosensor achieved 96% concordance with liquid chromatography-mass spectrometry/tandem mass spectrometry (LC-MS/MS) results on samples that underwent standard opiate testing (n=50). In addition, successful detection of a commonly abused opiate, oxycodone, resulted in 100% qualitative agreement between the PSi opiate sensor and LC-MS/MS. In contrast, a commercial broad opiate immunoassay technique (CEDIA®) achieved 65% qualitative concordance with LC-MS/MS. Evaluation of important performance attributes including precision, accuracy, and recovery was completed on blank urine specimens spiked with test analytes. Variability of morphine detection as a model opiate target was < 9% both within-run and between-day at and above the cutoff limit of 300 ng ml−1. This study validates the analytical screening capability of label-free PSi opiate immunosensors in authentic patient samples and is the first semi-quantitative demonstration of the technology’s successful clinical use. These results motivate future development of PSi technology to reduce complexity and cost of diagnostic testing particularly in a point-of-care setting.
DOI: 10.1373/clinchem.2009.127514
发表时间: 2010-02
期刊: Clinical chemistry
影响因子: 9.3
作者:
Ellington AA;Kullo IJ;Bailey KR;Klee GG
通讯作者: Klee GG
DOI: 10.1021/ac0488208
发表时间: 2004-12-01
影响因子: 7.4
作者:
DeLouise, LA;Miller, BL
通讯作者: Miller, BL
DOI: 10.1126/science.1081298
发表时间: 2003-03-28
期刊: SCIENCE
影响因子: 56.9
作者:
Li, YY;Cunin, F;Sailor, MJ
通讯作者: Sailor, MJ
DOI: 10.1021/ac902453h
发表时间: 2010-01-15
影响因子: 7.4
作者:
Bonanno, Lisa M.;DeLouise, Lisa A.
通讯作者: DeLouise, Lisa A.
DOI: 10.1126/science.278.5339.840
发表时间: 1997-10-31
期刊: SCIENCE
影响因子: 56.9
作者:
Lin, VSY;Motesharei, K;Ghadiri, MR
通讯作者: Ghadiri, MR