Algorithm for interpreting the results of frequency doubling perimetry.

Algorithm for interpreting the results of frequency doubling perimetry.
复制标题

用于解释倍频视野检查结果的算法。

DOI:
10.1016/s0002-9394(99)00399-2
复制
发表时间:
2000
影响因子:
4.2
通讯作者:
A. Robin
A. Robin
中科院分区:
医学1区
文献类型:
--
作者:
S. Patel;D. Friedman;Parna Varadkar;A. Robin

文献摘要

参考文献

被引文献

相似文献

目的评价一种基于倍频技术筛选模式的青光眼视野缺损识别算法。方法对150例连续青光眼患者中的137例进行倍频技术筛选模式和瑞典交互阈值算法视野检查。我们为倍频技术创建了一个算法,该算法对更严重的缺陷和更接近固定的缺陷都给予了更高的重视。然后将这些值与通过青光眼半视野测试评估的瑞典交互阈值算法视野的结果进行比较,两名蒙面青光眼专家,和已发表的青光眼损伤的定义,以确定用于检测青光眼的倍频技术筛查模式的灵敏度和特异性。或更高,特异性为93%或更高,与定义青光眼的标准无关。在检测中度或重度场丢失时,灵敏度增加到95%,而特异性不受影响。所有视野正常的受试者均在90秒内完成倍频技术检查。在大多数情况下,眼睛是异常的,如果他们有两个周边缺陷或一个中央default.CONCLUSIONSOUR倍频技术算法检测青光眼视野缺陷,允许在不到5分钟的双边筛选具有高灵敏度和高特异性。
PURPOSETo evaluate an algorithm for the identification of glaucomatous visual field defects with the screening mode of frequency doubling technology.METHODSScreening-mode frequency doubling technology and Swedish interactive threshold algorithm perimetry were performed on 137 of 150 consecutive patients referred to a glaucoma specialist. We created an algorithm for the frequency doubling technology that gave increased importance to both more severe defects and defects closer to fixation. These values were then compared with the results of the Swedish interactive threshold algorithm visual fields evaluated by the glaucoma hemifield test, two masked glaucoma specialists, and a published definition of glaucomatous damage to determine sensitivity and specificity of the frequency doubling technology screening mode for detecting glaucoma.RESULTSThe frequency doubling technology score obtained with our algorithm had a sensitivity of 80% or more and specificity of 93% or more irrespective of the criteria for defining glaucoma. Sensitivity increased to 95% in detecting moderate or severe field loss without a compromise in specificity. All subjects with normal visual fields completed the frequency doubling technology examination within 90 seconds. In most cases, eyes were abnormal if they had two peripheral defects or one central defect.CONCLUSIONSOur frequency doubling technology algorithm had both high sensitivity and high specificity in detecting glaucomatous visual field defects, allowing for bilateral screening in less than 5 minutes.
在医疗诊所使用神经纤维层照片筛查青光眼。
DOI: 10.1001/archopht.1994.01090180094042
发表时间: 1994
期刊: Archives of ophthalmology (Chicago, Ill. : 1960)
影响因子: --
作者:
Wang,F;Quigley,HA;Tielsch,JM
通讯作者: Tielsch,JM
DOI: 10.1136/bjo.80.5.389
发表时间: 1996-05-01
影响因子: 4.1
作者:
Quigley, HA
通讯作者: Quigley, HA
DOI: 10.1001/archopht.1995.01100050054031
发表时间: 1995-05-01
影响因子: --
作者:
SCHUMAN, JS;HEE, MR;FUJIMOTO, JG
通讯作者: FUJIMOTO, JG
DOI: 10.1056/nejm199505043321806
发表时间: 1995-05-04
影响因子: 158.5
作者:
TIELSCH, JM;JAVITT, JC;SOMMER, A
通讯作者: SOMMER, A
DOI: 10.1093/oxfordjournals.aje.a116013
发表时间: 1991-11-15
影响因子: 5
作者:
TIELSCH, JM;KATZ, J;SOMMER, A
通讯作者: SOMMER, A