ETD in a traveling wave ion guide at tuned Z-spray ion source conditions allows for site-specific hydrogen/deuterium exchange measurements.

ETD in a traveling wave ion guide at tuned Z-spray ion source conditions allows for site-specific hydrogen/deuterium exchange measurements.
复制标题

DOI:
10.1007/s13361-011-0196-7
复制
发表时间:
2011-10
影响因子:
3.2
通讯作者:
Brown, Jeffery M.
Brown, Jeffery M.
中科院分区:
化学3区
文献类型:
--
作者:
Rand, Kasper D.;Pringle, Steven D.;Morris, Michael;Engen, John R.;Brown, Jeffery M.

文献摘要

参考文献

被引文献

相似文献

最近应用电子转移解离(ETD)在单残基分辨率(HX-ETD)下测量溶液中蛋白质的氢交换,为基于质谱的生物分子结构分析铺平了道路,达到了前所未有的详细水平。该方法要求在ETD之前多肽离子的激活最小,以防止溶液中氘标签的不良气相随机化(即氢乱序)。在这里,我们探索了在四极飞行时间(Q-TOF)质谱仪的行波离子波导中使用“z喷雾”型离子源的ETD来测量肽中单个残基的氘含量。我们系统地确定了有助于碰撞激活的z喷射离子源的关键参数,并确定了允许在行波离子波导中进行无气相氢扰的ETD实验的条件。我们发现,在随后的行波离子波导中,ETD和补充碰撞激活可以改善低电荷肽中残基特异性氘含量的提取。我们的结果证明了这种方法的可行性,并说明了在配备行波离子波导的高分辨率Q-TOF仪器上进行HX-ETD实验的优势。有助于离子加热的z喷射离子源参数的确定同样与越来越多的质谱应用相关,这些应用也依赖于离子向气相的能量温和转移,例如通过结合气相离子-离子/离子中性反应或离子迁移谱法的天然质谱法分析生物分子结构。
The recent application of electron transfer dissociation (ETD) to measure the hydrogen exchange of proteins in solution at single-residue resolution (HX-ETD) paves the way for mass spectrometry-based analyses of biomolecular structure at an unprecedented level of detail. The approach requires that activation of polypeptide ions prior to ETD is minimal so as to prevent undesirable gas-phase randomization of the deuterium label from solution (i.e., hydrogen scrambling). Here we explore the use of ETD in a traveling wave ion guide of a quadrupole-time-of-flight (Q-TOF) mass spectrometer with a “Z-spray” type ion source, to measure the deuterium content of individual residues in peptides. We systematically identify key parameters of the Z-spray ion source that contribute to collisional activation and define conditions that allow ETD experiments to be performed in the traveling wave ion guide without gas-phase hydrogen scrambling. We show that ETD and supplemental collisional activation in a subsequent traveling wave ion guide allows for improved extraction of residue-specific deuterium contents in peptides with low charge. Our results demonstrate the feasibility, and illustrate the advantages of performing HX-ETD experiments on a high-resolution Q-TOF instrument equipped with traveling wave ion guides. Determination of parameters of the Z-spray ion source that contribute to ion heating are similarly pertinent to a growing number of MS applications that also rely on an energetically gentle transfer of ions into the gas-phase, such as the analysis of biomolecular structure by native mass spectrometry in combination with gas-phase ion-ion/ion-neutral reactions or ion mobility spectrometry.
DOI: 10.1021/ac901736j
发表时间: 2009-11-15
影响因子: 7.4
作者:
Nagy, Kornel;Redeuil, Karine;Rezzi, Serge
通讯作者: Rezzi, Serge
研究马细胞色素 c 蛋白氢交换的决定因素
DOI: 10.1002/pro.5560070323
发表时间: 1998-03-01
期刊: PROTEIN SCIENCE
影响因子: 8
作者:
Milne, JS;Mayne, L;Englander, SW
通讯作者: Englander, SW
DOI: 10.1021/ja904379w
发表时间: 2009-09-09
影响因子: 15
作者:
Pan, Jingxi;Han, Jun;Konermann, Lars
通讯作者: Konermann, Lars
DOI: 10.1021/ac901366n
发表时间: 2009-10-01
影响因子: 7.4
作者:
Kaltashov, Igor A.;Bobst, Cedric E.;Abzalimov, Rinat R.
通讯作者: Abzalimov, Rinat R.
DOI: 10.1038/nmeth.1265
发表时间: 2008-11-01
期刊: NATURE METHODS
影响因子: 48
作者:
Heck, Albert J. R.
通讯作者: Heck, Albert J. R.