MEMBRANE AUTOTRANSFUSION SYSTEM
MEMBRANE AUTOTRANSFUSION SYSTEM
批准号:
3501869
负责人:
UDIPI R SHETTIGAR
金额:
$5.0万
依托单位:
依托单位国家:
美国
项目类别:
财政年份:
1991
资助国家:
美国
项目状态:
已结题
起止时间:
1991-08-01 至 1992-01-31
中文摘要
点击翻译按钮获取中文摘要
英文摘要
Intraoperative autotransfusion system (IOAT) aspirates waste blood from
surgical wound site, purifies and returns it to the same patient. We
developed and evaluated in-vivo a new IOAT system which is designed to
purify the aspirated waste blood using hollow fiber membrane
ultrafilter/plasma filters instead of using centrifuges, and return the
purified whole blood to the patient in a closed loop continuous on line
operation within a few minutes without significant wastage of blood
cells and clotting factors.
Preliminary in-vivo evaluations of our Membrane Autotransfusion System
(MATS) during implantation of artificial ventricles in calves have lead
us to redefine the design parameters. We rebuilt our MATS so that it
could respond to the rapidly changing operating conditions, such as
blood flow rate (o to 2,000 ml/min, hematocrit (o to 45%), air emboli &
froth, macroaggregates, irrigating fluid, shear rate, transmembrane
pressure, duration of operation, and systemic heparinization.
Phase-I involves extensive in-vivo studies using calf and human blood
with an objective to establish the consistency of performance of the
system in response to the rapid changes in operating conditions. Phase-
II involves evaluations (i) during total artificial heart implantation
studies in calves, and (ii) in ten calves, where bleeding will be
simulated in open chest cavity, and the shed blood will be salvaged by
our membrane autotransfusion system. Total available market is expected
to be over $120,000,000 in the U.S. There is a strong need for this
device in view of the present scare in donor blood transfusions related
to transmission of AIDS and hepatitis.
期刊论文(0)
专著(0)
科研奖励(0)
会议论文
MEMBRANE AUTOTRANSFUSION SYSTEM (MATS)
-
批准号:3508786
-
项目类别:
-
资助金额:$9.0万
-
财政年份:1991
-
负责人:UDIPI R SHETTIGAR
-
依托单位: