LASER BASED MONOENERGETIC AND TUNABLE X RAY SYSTEM
LASER BASED MONOENERGETIC AND TUNABLE X RAY SYSTEM
批准号:
2750426
负责人:
Andrzej Krol
金额:
$19.79万
依托单位国家:
美国
项目类别:
财政年份:
1996
资助国家:
美国
项目状态:
已结题
起止时间:
1996-08-01 至 2000-07-31
中文摘要
点击翻译按钮获取中文摘要
英文摘要
DESCRIPTION (Adapted from Applicant's Abstract): The conventional x-ray
tube is characterized by low output, low brightness, a large focal spot and
a polychromatic photon energy distribution. The large focal spot problem is
especially problematic in magnification work. The low output mandates
undersirable long exposure times. The polychromatic spectral distribution,
with very limited mean photon energy control, impedes effective application
of DSA and makes DESA impractical. The applicants proposed a new x-ray
source that is created by focusing a high energy laser pulse on the surface
of solids. This laser-based source is very bright and delivers radiation in
the form of very short pulses (below 1 us) from a very small focal spot area
(as low as 10 um). The estimated cost of a clinical system is comparable to
the rotating anode-based units. An additional advantage of this new
laser-based source rests on the fact that it has a disposable target which
is replaced after each laser pulse. The target's elemental composition can
be thus rapidly (tens of Hz) changed between subsequent laser pulses. This
allows tailoring (including optimized K-edge filters) of the spectral
position of characteristic emission lines to the specific imaging tast (e.g.
patient thickness in magnification studies or the K-edge of a contrast agent
in DSA studies). Due to the high brightness of the laser-based x-ray
source, a spherically bent crystal monochromator can be inserted between the
target and the patient allowing utilization of near-monoenergetic x-rays for
true digital energy subtraction angiography studies. The polychromatic
structure of the x-ray beam generated by an electron-impact tube does not
allow precise matching to the K-edge of the object of interest and is always
mismatched with the optimal energy response of the typical x-ray detectors
used in radiology. Therefore, one could expect that the application of the
near-monoenergetic x-ray source will result in lowering the integral dose by
precisely tuning the x-ray energy according to the subject's composition and
thickness and by better matching the x-ray energy with the optimized
detector. In this proposed project, the applicants proposed to study,
develop, evaluate, and optimize a laser-based ultra-microfocused and very
short pulsed laser-based x-ray source for application in magnification and
DSA studies and monochromatic x-ray system LAMOX (LAser-based Monochromatic
X-rays) for application in digital energy subtraction angiography (DESA).
This project will be realized with close collaboration among x-ray
physicists (monochromator), laser physicists (laser x-ray source), medical
physicists, a physiologist and radiologists (evaluation and optimization of
the system through phantom studies). The specific goals of this project are
to optimize image quality and dose utilization and to demonstrate the
economic feasibility of the laser-based x-ray systems as a viable clinical,
hospital based, alternative to traditional magnification, DSA and DESA
systems.
期刊论文(3)
专著(0)
科研奖励(0)
会议论文
Ultrafast laser-based x-ray in-vivo phase-contrast micro-CT
-
批准号:7140668
-
项目类别:
-
资助金额:$17.13万
-
财政年份:2005
-
负责人:Andrzej Krol
-
依托单位:
Ultrafast laser x-ray in-vivo phase-contrast micro-CT
-
批准号:7035529
-
项目类别:
-
资助金额:$22.62万
-
财政年份:2005
-
负责人:Andrzej Krol
-
依托单位:
LASER BASED MONOENERGETIC AND TUNABLE X RAY SYSTEM
-
批准号:2460059
-
项目类别:
-
资助金额:$19.49万
-
财政年份:1996
-
负责人:Andrzej Krol
-
依托单位:
LASER BASED MONOENERGETIC AND TUNABLE X RAY SYSTEM
-
批准号:2230146
-
项目类别:
-
资助金额:$25.73万
-
财政年份:1996
-
负责人:Andrzej Krol
-
依托单位:
海外基金