Clinical evaluation of a new single-tube multiplex reverse transcription PCR assay for simultaneous detection of 11 respiratory viruses, Mycoplasma pneumoniae and Chlamydia in hospitalized children with acute respiratory infections.

Clinical evaluation of a new single-tube multiplex reverse transcription PCR assay for simultaneous detection of 11 respiratory viruses, Mycoplasma pneumoniae and Chlamydia in hospitalized children with acute respiratory infections.
复制标题

DOI:
10.1016/j.diagmicrobio.2017.03.010
复制
发表时间:
2017-06
影响因子:
2.9
通讯作者:
Ma XJ
Ma XJ
中科院分区:
医学4区
文献类型:
--
作者:
Zhao MC;Li GX;Zhang D;Zhou HY;Wang H;Yang S;Wang L;Feng ZS;Ma XJ

文献摘要

参考文献

被引文献

相似文献

呼吸道病原体13检测试剂盒(13×试剂盒)能够在单一反应中同时检测11种呼吸道病毒、肺炎支原体(MP)和衣原体。使用572例住院儿童鼻咽吸出物,评价13×试剂盒检测11种呼吸道病毒的临床性能,并与中国省级疾病预防控制中心常规使用的双管多重逆转录PCR检测(双管检测)进行比较。通过商业实时定量PCR(qPCR)试剂盒或测序评价13×试剂盒检测MP和衣原体的临床性能。对于供试病毒,试验一致性为95.98%,kappa系数为0.89。13×试剂盒检测MP和衣原体阳性标本均为真阳性。在临床环境中使用13×试剂盒将有助于医生根据病毒类型或多重感染评估临床结局,并限制抗生素的使用。13×试剂盒可同时检测MP和Ch等11种呼吸道病毒,检出率与双管法相当(kappa >0.75)。13×试剂盒的敏感性和特异性分别为97.41%和91.89%。 通过qPCR/测序验证13×试剂盒检测到的MP和衣原体为真阳性。
Respiratory Pathogen 13 Detection Kit (13× kit) is able to simultaneously detect 11 respiratory viruses, Mycoplasma pneumoniae (MP) and Chlamydia in a single reaction. Using 572 Nasopharyngeal aspirates collected from hospitalized children, the clinical performance of 13× kit for detecting 11 respiratory viruses was evaluated in comparison with a routinely used 2-tube multiplex reverse transcription PCR assay (2-tube assay) at provincial Centers for Disease Control and Prevention in China. The clinical performance of 13× kit for detecting MP and Chlamydia was evaluated by commercial real-time quantitative PCR (qPCR) kits or sequencing. For tested viruses, the assay concordance was 95.98% and the kappa coefficient was 0.89. All the MP and Chlamydia positive samples detected by 13× kit were confirmed as true positives. The utilization of the 13× kit in clinical settings will be helpful for doctors to assess clinical outcome according to virus type or multiple infections, and to limit the use of antibiotics. A 13× kit allowed for simultaneous detection of 11 respiratory viruses, MP and Ch. The detection rate of 13× kit was comparable to that of 2-tube assay (kappa >0.75). A 13× kit has high sensitivity (97.41%) and specificity (91.89%). MP and Chlamydia detected by 13× kit were verified by qPCR/sequencing as true positives.
DOI: 10.1002/jmv.20008
发表时间: 2004-03-01
影响因子: 12.7
作者:
Coiras, MT;Aguilar, JC;Pérez-Breña, P
通讯作者: Pérez-Breña, P
DOI: 10.1128/jcm.02436-06
发表时间: 2007-09-01
影响因子: 9.4
作者:
Mahony, J.;Chong, S.;Janeczko, R.
通讯作者: Janeczko, R.
DOI: 10.1128/jcm.01297-08
发表时间: 2009-03-01
影响因子: 9.4
作者:
Raymond, Frederic;Carbonneau, Julie;Corbeil, Jacques
通讯作者: Corbeil, Jacques
DOI: 10.1016/j.jviromet.2008.11.007
发表时间: 2009-03
影响因子: 3.1
作者:
Kim SR;Ki CS;Lee NY
通讯作者: Lee NY
DOI: 10.1186/s12879-015-0863-6
发表时间: 2015-03-13
影响因子: 3.7
作者:
Cui B;Zhang D;Pan H;Zhang F;Farrar J;Law F;van Doorn HR;Wu B;Ba-Thein W
通讯作者: Ba-Thein W