Development of a System Containing a Flow-Through Electrochemiluminescent Detector and Its Application to High-Sensitive Determination

Development of a System Containing a Flow-Through Electrochemiluminescent Detector and Its Application to High-Sensitive Determination
流通式电化学发光检测器系统的研制及其在高灵敏测定中的应用
批准号:
11650835
负责人:
KUSU Fumiyo
金额:
$2.11万
依托单位国家:
日本
项目类别:
Grant-in-Aid for Scientific Research (C)
财政年份:
1999
资助国家:
日本
项目状态:
已结题
起止时间:
1999 至 2000

项目摘要

项目成果

KUSU Fumiyo的其他基金

相似基金

相关文献

中文摘要
翻译
建立了一种用于生物活性物质高灵敏测定的流动电致发光检测器系统。主要研究内容如下:1.电致发光检测系统的构建该系统由载液池、脱气器、泵、试液进样器、电致发光池和光电管组成。对实验条件如载液组成、流动线所用聚四氟乙烯管的尺寸和长度以及池的设计进行了优化,得到了重复性好、灵敏度高的数据.高灵敏度测定过氧化氢采用鲁米诺电致发光法测定过氧化氢。使用5.0 μ M鲁米诺的磷酸盐缓冲液(pH 7.6),流速为0.4 ml/min,外加电位为1.1 V(相对于Ag/AgCl),可测定59 fmol至59 nmol范围内的过氧化氢量。3.血清分析作为生物样品分析的实例,测定血清中的葡萄糖和尿酸。将葡萄糖氧化酶或尿酸酶固定在填充在位于进样器和检测器之间的酶反应管中的壳聚糖上。本系统应适用于一个简单的方法,用于确定具有高灵敏度的生物重要物质。
英文摘要
A system containing a flow-through electrochemiluminescent detector was developed for high-sensitive determination of biologically important substances. The main contents of the results are as follows :1. Construction of the electrochemiluminescent detection systemThe system developed was fabricated from a carrier solution reservoir, degasser, pump, injector of test solution, electrochemiluminescent cell and phototube. The experimental conditions, such as the composition of carrier solution, the size and length of Teflon tube for the flow line and the cell design, were optimized to give reproducible data with high sensitivity.2. High-sensitive determination of hydrogen peroxideDetermination of hydrogen peroxide was carried out using luminol, an electrochemiluminescent reagent. Hydrogen peroxide amount ranging from 59 fmol to 59 nmol could be determined using 5.0 μM luminol in a phosphate buffer (pH 7.6) at a flow rate of 0.4 ml/min and an applied potential of 1.1 V vs. Ag/AgCl.3. Analysis of serumAs examples of analysis of biological samples, glucose and uric acid in serum were determined. Glucose oxidase or uricase were immobilized on Chitopearl filled in an enzyme reactor tube located between the injector and detector. The present system should be applicable to a simple method for determining biologically important substances with high sensitivity.
期刊论文(4)
专著(0)
科研奖励(0)
会议论文
DOI: --
发表时间:
期刊:
影响因子: --
作者: []
通讯作者:
DOI: --
发表时间:
期刊:
影响因子: --
作者: []
通讯作者:
DOI: --
发表时间:
期刊:
影响因子: --
作者: []
通讯作者:
Method Development for Assessing the Progression of Metabolic Syndrome by Multiplex-dynamic Analysis of Compounds of Biological Importance
Development of a System for Determining a Small Amount of Acid Based on Electrolysis of Quinone
Development of Flow-cell System with a Column Electrode for Various Analytical Uses
In situ Observation of Trace Amounts of Adsorbed species at Alloy Electrode/Solution Interface
海外基金