Biocompatibility evaluation of a thermoplastic rubber for wireless telemetric intracranial pressure sensor coating.
Biocompatibility evaluation of a thermoplastic rubber for wireless telemetric intracranial pressure sensor coating.
复制标题
DOI:
10.1016/j.jmbbm.2015.01.018
复制
发表时间:
2015-05
影响因子:
3.9
通讯作者:
Luciano, Mark G.
中科院分区:
文献类型:
--
作者:
Yang, Jun;Charif, Andrea C.;Puskas, Judit E.;Phillips, Hannah;Shanahan, Kaitlyn J.;Garsed, Jessica;Fleischman, Aaron;Goldman, Ken;Luebbers, Matthew T.;Dombrowski, Stephen M.;Luciano, Mark G.
This study investigated the biocompatibility of the experimental thermoplastic rubber Arbomatrix™ that will be used as the protective coating on a novel intracranial pressure (ICP) sensor silicon chip. Arbomatrix™ was benchmarked against biocompatible commercial silicone rubber shunt tubing in the brain via a rat model with 60-day implant duration. A bare silicon chip was also implanted. The results showed similar cellular distribution in the brain-implant boundary and surrounding tissues. Quantitative analysis of neuron and glia density did not show significant difference between implants. Through histological and immunohistochemical evaluation we conclude that Arbomatrix™ is well tolerated by the brain. Due to its exceptional barrier properties Arbomatrix™ has already been shown to be an excellent protective coating for new ICP monitoring chip.
登录
查看更多内容
影响因子:
12.7
作者:
STENSAAS, SS;STENSAAS, LJ
通讯作者:
STENSAAS, LJ
影响因子:
3.3
作者:
ENG, LF
通讯作者:
ENG, LF
影响因子:
2.4
作者:
AMMAR, A
通讯作者:
AMMAR, A
影响因子:
4.1
作者:
DYMOND, AM;KAECHELE, LE;CRANDALL, PH
通讯作者:
CRANDALL, PH
影响因子:
4.2
作者:
WEST, MJ
通讯作者:
WEST, MJ