A phantom tissue system for the calibration of perfusion measurements.
A phantom tissue system for the calibration of perfusion measurements.
复制标题
用于校准灌注测量的模型组织系统。
DOI:
10.1115/1.2948417
复制
发表时间:
2008
期刊:
影响因子:
--
通讯作者:
Diller,ThomasE
中科院分区:
文献类型:
--
作者:
Mudaliar,AshvinikumarV;Ellis,BrentE;Ricketts,PatriciaL;Lanz,OttoI;Scott,ElaineP;Diller,ThomasE
A convenient method for testing and calibrating surface perfusion sensors has been developed. A phantom tissue model is used to simulate the nondirectional blood flow of tissue perfusion. A computational fluid dynamics (CFD) model was constructed in Fluent® to design the phantom tissue and validate the experimental results. The phantom perfusion system was used with a perfusion sensor based on clearance of thermal energy. A heat flux gage measures the heat flux response of tissue when a thermal event (convective cooling) is applied. The blood perfusion and contact resistance are estimated by a parameter estimation code. From the experimental and analytical results, it was concluded that the probe displayed good measurement repeatability and sensitivity. The experimental perfusion measurements in the tissue were in good agreement with those of the CFD models and demonstrated the value of the phantom tissue system.