OPTIMIZATION OF A MICROFLUIDIC DEVICE FOR DIFFUSION-BASED EXTRACTION OF DMSO FROM A CELL SUSPENSION.
OPTIMIZATION OF A MICROFLUIDIC DEVICE FOR DIFFUSION-BASED EXTRACTION OF DMSO FROM A CELL SUSPENSION.
复制标题
DOI:
10.1016/j.ijheatmasstransfer.2008.04.018
复制
发表时间:
2008-11
影响因子:
5.2
通讯作者:
Hubel, A.
中科院分区:
文献类型:
--
作者:
Glass, K. K. Fleming;Longmire, E. K.;Hubel, A.
This study considers the use of a two-stream microfluidic device for extraction of dimethyl sulphoxide (DMSO) from a cryopreserved cell suspension. The DMSO diffuses from a cell suspension stream into a neighboring wash stream flowing in parallel. The model of Fleming et al. is employed to determine and discuss optimal geometry and operating conditions for a case requiring removal of 95% DMSO from suspension streams with volumetric flow rates up to 2.5 ml/min. The effects of Peclet number, flow rate fraction, and cell volume fraction are analyzed, and expansion of the analysis to other applications is discussed.