中文摘要
点击翻译按钮获取中文摘要
英文摘要
During the previous two years of effort, we have demonstrated that the Wien
mass spectrometer (WMS) is capable of analyzing compounds of very high mass
((CsI)155Cs+ at 40,433d) and with very high sensitivity (1 pmol of Bovine
insulin at 5,734d). The short analysis time of about 10 min/sample allows
rapid optimization of instrumental operating conditions such as sample
matrix composition and matrix temperature. Our primary goals will be to
establish the upper mass and sensitivity capability of the LSIMS technique
on the WMS and to collaborate with researchers at UCSF and elsewhere on a
wide variety of problems in biochemical engineering and structural
biochemistry. In support of these goals, several major improvements in the
WMS are proposed:
1) Install a more powerful Wien analyzer capable of increasing mass
resolution by a factor of four, without reducing sensitivity. This will
increase the accuracy of mass measurement by a commensurate amount;
2) Reduce instrument background pressure by increasing pumping speed. This
will improve transmission and resolution for high mass compounds;
3) Provide the instrumental capability for the analysis of negatively
charged compounds;
4) Continue refinements on the ion source, probe and detector to increase
sensitivity of the instrument.
期刊论文(2)
专著(0)
科研奖励(0)
会议论文
Effect of primary beam energy on the secondary ion sputtering efficiency of liquid secondary ionization mass spectrometry in the 5-30-keV range.
一次束能量对 5-30 keV 范围内液体二次电离质谱二次离子溅射效率的影响。
DOI:
10.1021/ac00165a016
发表时间:
1988
期刊:
Analytical chemistry
影响因子:
7.4
作者:
[Aberth,WH, Burlingame,AL]
通讯作者:
Burlingame,AL
Instrumental conditions of secondary ion mass spectrometry that affect sensitivity for observation of very high masses.
影响极高质量观测灵敏度的二次离子质谱仪条件。
DOI:
10.1021/ac00297a055
发表时间:
1986
期刊:
Analytical chemistry
影响因子:
7.4
作者:
[Aberth,W]
通讯作者:
Aberth,W