Closed-loop subcutaneous insulin infusion algorithm with a short-acting insuin analog for long-term clinical application of a wearable artificial endocrine pancreas.

Closed-loop subcutaneous insulin infusion algorithm with a short-acting insuin analog for long-term clinical application of a wearable artificial endocrine pancreas.
采用短效胰岛素类似物的闭环皮下胰岛素输注算法,用于可穿戴人工内分泌胰腺的长期临床应用。
批准号:
08680930
负责人:
SAKAKIDA Miciharu
金额:
$1.6万
依托单位:
依托单位国家:
日本
项目类别:
Grant-in-Aid for Scientific Research (C)
财政年份:
1996
资助国家:
日本
项目状态:
已结题
起止时间:
1996 至 1997

项目摘要

项目成果

相似基金

相关文献

中文摘要
翻译
点击翻译按钮获取中文摘要
英文摘要
Considering the management and safety of the insulin delivery route when a wearable artificial endocrine pancreas is applied to ambulatory diabetic patients for long-term basis, we developed a subcutaneous insulin infusion algorithm by analyzing the dynamics of subcutaneously injected short-acting insulin analogue (Insulin Lispro) by a three-compartmental model. Principally, the insulin infusion algorithm was developed as a transfer function with the first order delay in both proportional and derivative actions to blood glucose concentrations. The parameters for this algorithm were calculated to simulate a physiological plasma insulin profile as possible.By applying this algorithm with regular insulin, diabetic patients showed a 2-hr postprandial hyperglycemia and a delayd hyperinsulinemia, folowed by hypoglycemic episodes during 4 to 5 hr after oral glucose load, just as obervrd in computer simulation study. However using Insulin Lispro, a near-physiological glycemic control (postprandial blood glucose of 153.1(]SY.+-。[)8.3 mg/100 ml at 60 min, and 90.3(]SY.+-。[)7.1 mg/100 ml at 180 min, respectively) could be achieved without showing any delayd hyperinsulinemia nor hypoglycemia. Daily glycemic excursions were also controlled near-physiolosically and though the daily insulin requirement (731.7(]SY.+-。[)160.5 mU/kg/day) was slightly higher, but not significantly different from that with intravenous insulin infusion (622.3(]SY.+-。[)142.6 mU/kg/day).These results indicate that the application of the subcutaneous insulin infusion algorithm with Insulin Lispro is feasible for long-term glycemic control with the wearable artificial endocrine pancreas in ambulatory diabetic patients.
期刊论文(0)
专著(0)
科研奖励(0)
会议论文
S.Shimoda, K.Nishida, M.Sakakida et al.: "The clinical application of closed-loop subcutaneous insulin infusion algorithm with short-acting insulin analogue for wearable artificial endocrine pancreas." Jpn J Artif Organs. 26(3). 739-743 (1997)
S.Shimoda、K.Nishida、M.Sakakida 等人:“短效胰岛素类似物闭环皮下胰岛素输注算法在可穿戴人工内分泌胰腺中的临床应用”。
DOI: --
发表时间:
期刊:
影响因子: --
作者: []
通讯作者:
七里元亮ら: "人工膵島の現況と課題" 総合臨床. 46(11). 2809-2815 (1997)
Motosuke Shichiri 等人:“人工胰岛的现状和问题”《一般临床实践》46(11) (1997)。
DOI: --
发表时间:
期刊:
影响因子: --
作者: []
通讯作者:
S.Shimoda, K.Nishida, M.Sakakida et al.: "Closed-loop subcutaneous insulin infusion algorithm with a short-acting insulin analogue for long-term clinical application of a wearable artificial endocrine pancreas." Frontiers Med.Biol.Engng. 8(3). 197-211 (19
S.Shimoda、K.Nishida、M.Sakakida 等人:“采用短效胰岛素类似物的闭环皮下胰岛素输注算法,用于可穿戴人工内分泌胰腺的长期临床应用。”
DOI: --
发表时间:
期刊:
影响因子: --
作者: []
通讯作者:
18
    海外基金