课题基金 / 基金详情

EVALUATION OF ANALYTICAL COUNTERCURRENT CHROMATOGRAPHS

EVALUATION OF ANALYTICAL COUNTERCURRENT CHROMATOGRAPHS
分析逆流色谱的评估
批准号:
3878937
负责人:
Y ITO
金额:
$0.0万
依托单位国家:
美国
项目类别:
财政年份:
--
资助国家:
美国
项目状态:
未结题
起止时间:

项目摘要

项目成果

Y ITO的其他基金

相似基金

相关文献

中文摘要
翻译
点击翻译按钮获取中文摘要
英文摘要
Performance of two countercurrent chromatographic models, high-speed countercurrent chromatograph (HSCCC-4000) and analytical toroidal coil centrifuge (TCC), is evaluated in terms of theoretical plate number, resolution factor and separation time to assess their analytical capability. A series of experiments was conducted to investigate the effects of internal diameter and length of the coiled column and flow rate of the mobile phase on the separation of indole auxins in two-phase solvent systems composed of n-hexane/ethyl acetate/methanol/water at different volume ratios. The three components of the indole auxins were completely resolved in 16 min with the HSCCC system equipped with a multilayer coil of a 0.55 mm ID PTFE tube with theoretical plates ranging from 829 to 1,290. Similar separation was achieved in 24 min with the TCC system equipped with a 0.3mm ID PTFE tube with theoretical plates ranging from 969 to 1,811. It is concluded that both systems have comparable analytical capability at the present stage of development.